A simple, selective, rapid, precise and economical reverse phase HPLC method has been developed for the L-Arginine effervescent powder. The method was carried out on a
Inertsil C8 (250 x 4.6 mm x 5 μm)) column with a mobile phase consisting of Solution A: Acetonitrile (920:80) where solution A contain 500 mg of octane sulphonic acid in 1000 ml of phosphate buffer (pH 3.5)
at a flow rate of 1.0 ml/min. Detection was carried out at 215 nm. The retention time of L-Arginine was found to be 7.5 minutes. The developed method was validated in terms of system suitability, linearity, accuracy& %recovery, precision, robustness, specificity, intermediate precision.
The proposed method can be used for the estimation of L-Arginine.
The proposed method is optimized and validated as per ICH guidelines.